2023 July: Difference between revisions
Line 6: | Line 6: | ||
== July 22 == | == July 22 == | ||
'''00:10 UT''' The new feed + radome was installed on Ant 11 and the control files were changed to add Ant 11 tracking. The pointing measurement at 21:30 UT shows that Ant 11 is not pointing well in HA, so this needs to be updated. I will take more measurements tomorrow and adjust the pointing. The phase calibration at 23:14 UT showed coherence at low frequencies on Ant 11, so I was able to adjust the delay (about 3 ns) to optimize that. We have a weekend coming up, so we won't be able to explore focus adjustments, but I will do a series of observational tests with this current focus over the weekend. | '''00:10 UT''' The new feed + radome was installed on Ant 11 and the control files were changed to add Ant 11 tracking. The pointing measurement at 21:30 UT shows that Ant 11 is not pointing well in HA, so this needs to be updated. I will take more measurements tomorrow and adjust the pointing. The phase calibration at 23:14 UT showed coherence at low frequencies on Ant 11, so I was able to adjust the delay (about 3 ns) to optimize that. We have a weekend coming up, so we won't be able to explore focus adjustments, but I will do a series of observational tests with this current focus over the weekend. | ||
== July 24 == | |||
'''17:23 UT''' The feed orientation was off by 45 degrees. Owen traced that to having the plate installed backwards, so it is now flipped and the orientation should be right. I am doing a SOLPNTCAL to check it. |
Revision as of 17:24, 24 July 2023
July 16
23:00 UT In the afternoon all of the Laird temperature controllers stopped regulating, allowing the FEMs to reach >40 C. This was likely heat-related due to unusually high ambient temperatures. Things were back to normal by the next day.
July 22
00:10 UT The new feed + radome was installed on Ant 11 and the control files were changed to add Ant 11 tracking. The pointing measurement at 21:30 UT shows that Ant 11 is not pointing well in HA, so this needs to be updated. I will take more measurements tomorrow and adjust the pointing. The phase calibration at 23:14 UT showed coherence at low frequencies on Ant 11, so I was able to adjust the delay (about 3 ns) to optimize that. We have a weekend coming up, so we won't be able to explore focus adjustments, but I will do a series of observational tests with this current focus over the weekend.
July 24
17:23 UT The feed orientation was off by 45 degrees. Owen traced that to having the plate installed backwards, so it is now flipped and the orientation should be right. I am doing a SOLPNTCAL to check it.